定体积位移增量法测量医用微小流量技术研究
引用本文:廖旭辉,陈昱桥,杨勇,王尧君,张秋辰,冉文杰,杨俊杰,李鹏.定体积位移增量法测量医用微小流量技术研究[J].计量学报,2020,41(8):937-941.
作者姓名:廖旭辉  陈昱桥  杨勇  王尧君  张秋辰  冉文杰  杨俊杰  李鹏
作者单位:1. 中国测试技术研究院, 四川 成都 610021
2. 乐山市计量测试所, 四川 乐山 614000
3. 四川中测辐射科技有限公司, 四川 成都 610021
摘    要:利用定体积活塞腔体测量微流量原理,在建立活塞位移量与流速的关系基础上,研究了医用微小流量的测量模型、理论计算方法、硬件设计以及对应的软件系统,该微流量测量技术能在短时间内完成对医用输液泵和注射泵的校准工作。实验结果表明,流量检测装置可测量1~1500mL/h的微小流量,不仅能够快速响应1mL/h微小流量,且测量瞬时流量值最大允许误差不大于±1%;当注射容量腔体液体体积大于5mL时所测得累计流量最大允许误差不大于±1%。

关 键 词:计量学  微小流量  定型腔  位移增量  推杆  
收稿时间:2019-01-07

Research on Measurement of Medical Micro Flow Rate Technology Based on Constant Volume Displacement Increment Method
LIAO Xu-hui,CHEN Yu-qiao,YANG Yong,WANG Yao-jun,ZHANG Qiu-chen,RAN Wen-jie,YANG Jun-jie,LI Peng.Research on Measurement of Medical Micro Flow Rate Technology Based on Constant Volume Displacement Increment Method[J].Acta Metrologica Sinica,2020,41(8):937-941.
Authors:LIAO Xu-hui  CHEN Yu-qiao  YANG Yong  WANG Yao-jun  ZHANG Qiu-chen  RAN Wen-jie  YANG Jun-jie  LI Peng
Affiliation:1. National Institute of Measurement and Testing Technology, Chengdu, Sichuan 610021, China
2. Leshan Metrological Measurement Institute, Leshan, Sichuan 614000, China
3. Sichuan Zhongce Radiation Technology Co. Ltd, Chengdu, Sichuan 610021, China
Abstract:Using the principle of measuring the micro-flow rate of a fixed-volume piston cavity, based on the relationship between the piston displacement and the flow rate, the measurement model, theoretical calculation method, hardware design and corresponding software system of medical micro-flow are introduced. The micro-flow measurement technology can complete the calibration of medical infusion pump and injection pump in a short time. The experimental results show that the flow detector can measure the micro-flow rate from 1mL/h to 1500mL/h, and can not only respond to the micro-flow rate of 1mL/h quickly, but also the maximum allowable error of instantaneous flow rate is less than ±1%. The maximum allowable error of cumulative flow rate is less than ±1% when the volume of liquid in the injection chamber is larger than 5mL.
Keywords:metrology  microflow  fixed volume cavity  displacement increment  push rod  