5Hz~10kHz频率范围矢量水听器国际主导比对
引用本文:陈毅,贾广慧,费腾,Isaev A E,Matveev A N.5Hz~10kHz频率范围矢量水听器国际主导比对[J].计量学报,2020,41(10):1279-1283.
作者姓名:陈毅  贾广慧  费腾  Isaev A E  Matveev A N
作者单位:1.杭州应用声学研究所,浙江 杭州 311400
2.浙江大学 信息与电子工程学院,浙江 杭州 310027
3.俄罗斯物理技术与无线电测量工程研究院,俄罗斯 莫斯科 门德列沃 141570
摘    要:利用VHS56、VHS90和KPG-10作为比对传递用矢量水听器,中国杭州应用声学研究所与俄罗斯VNIIFTRI开展了 5Hz~10kHz 频率范围矢量水听器校准国际比对。利用各自的自由场互易法和垂直驻波管比较法校准装置,校准了3只矢量水听器X、Y、Z通道的灵敏度。从校准结果看,一致性较好,174个频率点中,仅有一个频点的校准值与参考值的偏差略微超过参考扩展不确定度,绝大部分频率点的测量偏差都小于±0.5dB,检验和验证了中俄双方矢量水听器校准装置的准确性和可靠性。

关 键 词:计量学  矢量水听器  灵敏度  国际比对  
收稿时间:2019-10-10

The Pilot Comparison Calibration of Vector Receivers in the Frequency Range 5 Hz to 10 kHz
CHEN Yi,JIA Guang-hui,FEI Teng,Isaev A E,Matveev A N.The Pilot Comparison Calibration of Vector Receivers in the Frequency Range 5 Hz to 10 kHz[J].Acta Metrologica Sinica,2020,41(10):1279-1283.
Authors:CHEN Yi  JIA Guang-hui  FEI Teng  Isaev A E  Matveev A N
Affiliation:1. Hangzhou Applied Acoustics Research Institute, Hangzhou, Zhejiang 311400, China
2. College of Information Science & Electronic Engineering, Zhejiang University, Hangzhou, Zhejiang 310027, China
3. Russian National Research Institute for Physicotechnical and Radio Engineering Measurements, Mendeleevo, Moscow Region 141570, Russia
Abstract:Using three vector receivers VHS56, VHS90 and KPG-10 as reference vector receivers, the comparison calibration of vector receivers in the frequency range 5Hz to 10kHz was performed between HAARI, China and VNIIFTRI, Russia . The sensitivities of X, Y and Z channels of three vector receivers were calibrated by free-field reciprocity calibration facilities and vertical standing wave tube comparison calibration facilities of HAARI and VNIIFTRI. The calibration results of two participants were consonant with each other. There was only one frequency points deviation between calibration results and reference values slightly exceeding the reference expand uncertainties in all 174 calibration frequency points, and the deviations at most frequency points were less than ±0.5dB, which tested and verified the correction and reliability of the calibration results of vector receivers given by both participants international calibration facilities.
Keywords:metrology  vector receiver  sensitivity  international comparison calibration  