首页 | 本学科首页   官方微博 | 高级检索  
     

BHP6602型小型γ相机的临床验证研究
引用本文:刘轶敏,李方,陈黎波,巴建涛,吕京桥.BHP6602型小型γ相机的临床验证研究[J].现代仪器,2004,10(4):49-51.
作者姓名:刘轶敏  李方  陈黎波  巴建涛  吕京桥
作者单位:北京协和医院核医学科,北京,100730
摘    要:甲状腺模型和 5 9例临床志愿者分别用BHP6 6 0 2型γ相机和Sigma 4 2 0γ相机进行甲状腺显像 (5 5例 )和甲状旁腺显像 (4例 ) ,并对显像结果进行比较。结果表明BHP6 6 0 2型γ相机与Sigma 4 2 0γ相机的显像结果与临床诊断符合率完全一致 ,前者图像质量优于后者。BHP6 6 0 2型γ相机完全适用于小脏器显像。

关 键 词:γ相机  质控测试  甲状腺  甲状旁腺

A clinical verification study of BHP6602γcamera
Liu Yimin Li Fang Chen Libo Ba Jiantao Lv Jingqiao.A clinical verification study of BHP6602γcamera[J].Modern Instruments,2004,10(4):49-51.
Authors:Liu Yimin Li Fang Chen Libo Ba Jiantao Lv Jingqiao
Affiliation:PUMC Hospital Beijing 100730
Abstract:A new domesticγcamera (Model : BHP6602) was tested for scintigraphing with phantom and patients, and it was compared with an imported γcamera (Sigma420, Technicare Corp.). The images were analyzed by two experienced physicians of nuclear medicine. The results showed that both machine hade the same lesion detectability, while the imaging quality of BHP6602 was better than that of Sigma420. This confirmed that BHP6602γcamera was qualified for clinical use.
Keywords:γcamera Quality control Thyroid Parathyroid
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号