Abstract:A digital microfluidic system based on electrowetting has been developed to facilitate the investigation of pre-steady-state reaction kinetics using rapid quenching and matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS). The device consists of individually addressable electrodes arranged to allow the combination of liquid droplets at well-defined time intervals and an integrated, electrohydrodynamically driven mixer. The device combines two droplets to initiate a reaction, then, with precise timing, combines a third droplet to quench the reaction, and finally combines a fourth droplet to form a matrix. Improvements to throughput when compared to traditional laboratory-scale methods, and previous MALDI-TOF MS digital microfluidic systems, were made. The device was tested against a model protein tyrosine phosphatase system, and results agreed well with published data. The system therefore allows for the analysis of reaction kinetics that were previously too rapid to analyze using MALDI-TOF MS.