首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of glucose in whole blood samples by mid-infrared spectroscopy
Authors:Kim Yoen-Joo  Hahn Sangjoon  Yoon Gilwon
Affiliation:Medical Application Team, Samsung Advanced Institute of Technology, P.O. Box 111, Suwon 440-600, Korea.
Abstract:We have determined the glucose concentration of whole blood from mid-infrared spectra without sample preparation or use of chemical reagents. We selected 1119-1022 cm(-1) as the optimal wavelength range for our measurement by making a first-loading vector analysis based on partial least-squares regression. We examined the influence of hemoglobin on samples by using different calibration and prediction sets. The accuracy of glucose prediction depended on the hemoglobin level in the calibration model; the sample set should represent the entire range of hemoglobin concentration. We obtained an accuracy of 5.9% in glucose prediction, and this value is well within a clinically acceptable range.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号