Novel design of a bioelectric amplifier with minimized magnitude and phase errors |
| |
Authors: | Mashhour Bani Amer |
| |
Affiliation: | Dept. of Biomedical Engineering, Faculty of Engineering,Jordan University of Science and Technology, P. O. Box 3030 Irbid 22110, JORDAN |
| |
Abstract: | A new design of a bioelectric amplifier that has better parameters than conventional designs is presented. The design allows the construction of bioelectric amplifier with improved parameters in terms of common-mode rejection ratio and phase and magnitude errors. The voltage gain is easily adapted to a wide range of biomedical applications. The experimental and simulation results of the designed bioelectric amplifier are also included. |
| |
Keywords: | Bioelectric amplifier Instrumentation amplifier Common-mode rejection ratio |
本文献已被 CNKI 万方数据 SpringerLink 等数据库收录! |