Construction and evaluation of an automated flow injection-stopped flow analyser for multipoint reaction rate spectrophotometric methods. Determination of ammonia nitrogen,creatinine and phosphate
Authors:Constantinos A. Georgiou  Michael A. Koupparis
Affiliation:. Laboratory of Analytical Chemistry, Department of Chemistry, University of Athens, Panepistimiopolis, Kouponia, Athens, 15771, Greece,
Abstract:The construction and evaluation of a fully automated Flow Injection-Stopped Flow (FI-SF) spectrophotometric analyser is described. A microcomputer (Rockwell AIM 65) is used to control the analyser (sample injection, stop and start of the pump) through a suitable interface. Data acquisition is achieved using a 12 bit ADC card and a suitable subroutine in 6502 assembly language, allowing data sampling at a frequency of 7.5 kHz. The measurement interface and software were evaluated using a voltage ramp generator. A precision of 0.02-1.1% RSD (N =10) was obtained for voltage ramps in the range of 1-37 mVs-1. The FI-SF analyser was evaluated in routine analysis by developing FI-SF kinetic spectrophotometric methods for the determination of ammonia nitrogen (20-250 ppm, 0.4-2.5% RSD) based on theBerthelot reaction, creatinine (20-220 ppm, 0.9-3.6% RSD) based on the Jaffé reaction, and phosphate (5-30 ppm, 1.0-3.3% RSD) based on the phosphomolybdenum blue reaction. The reaction rate is measured by linear fitting of multiple absorbance readings vs time. Algorithms for automated estimation of the residence time, the linear range of the reaction curve, and data treatment are presented.