Model‐Based Recognition of Mid‐Infrared Sensor Fouling in Paracetamol Crystallization
Authors:Clemens B. Minnich  Lydia Helmdach  Joachim Ulrich  Martin P. Feth
Affiliation:1. S‐PACT GmbH, Aachen, Germany.;2. Martin‐Luther‐Universit?t Halle‐Wittenberg, Zentrum für Ingenieurwissenschaften, Verfahrenstechnik/TVT, Halle, Germany.;3. Sanofi‐Aventis Deutschland GmbH, Chemical and Biotechnological Development (C&BD), Frankfurt Chemistry, Frankfurt a. Main, Germany.
Abstract:A procedure for the recognition of undesired solid deposition on the optical element of an attenuated total reflectance probe for online mid‐infrared spectroscopy during crystallization processes is presented. Unlike other approaches, the method exclusively uses intelligent spectral analysis for solid recognition and does not require additional devices. Indirect hard modeling enables the separation of the solid material's contribution to the mixture spectra from those contributions of the dissolved material and the solvents, expressed through the predicted component weights. The detection of the solid component weight is promising for implementation into process monitoring or control.
Keywords:Crystallization  Online mid‐infrared spectroscopy  Attenuated total reflectance  Probe fouling  Indirect hard modeling