Hearing impairment fitting: study of an auditory prosthesis with 7 entirely programmable filters |
| |
Authors: | CH Chouard |
| |
Affiliation: | Service ORL, H?pital Saint-Antoine, Paris. |
| |
Abstract: | We designed a 7-filter digital auditory prosthesis table prototype. For each of these filters frequency band width, amplification and compression were programmable in order to adapt these parameters to the deaf patient's audiometric particularities. We compared the hearing improvement it was possible to obtain with our prototype as a function of the number of filters (3, 4 or 7) and their frequency band width programmability. We tested 21 patients suffering from middle or severe neurosensory hearing loss. This study allows to demonstrate that a 7 programmable-width-filter strategy seems to be the most appropriate. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|