Compact double focusing mass spectrometer design for partial pressure measurements
Authors:D Ioanoviciu  C Cuna
Affiliation:Institute for Stable Isotopes, Cluj, P.O. Box 75, Romania
Abstract:A double focusing mass spectrometer with Wien filter and magnetic deflector (both located in the same pole-gap) is proposed as a partial pressure analyzer. The magnetic field area to accommodate a beam of given radial angular aperture and the partial pressure sensitivity are compared with those of the single focusing instruments. The numerical examples given show a mass resolution increase of more than two compared with a single focusing instrument.