首页 | 本学科首页   官方微博 | 高级检索  
     


Accuracy of the spring constant of atomic force microscopy cantilevers by finite element method
Authors:Chen Bo-Yi  Yeh Meng-Kao  Tai Nyan-Hwa
Affiliation:Department of Power Mechanical Engineering and Department of Materials Science and Engineering, National Tsing Hua University, Hsinchu 30013 Taiwan, Republic of China.
Abstract:Atomic force microscopy (AFM) probe with different functions can be used to measure the bonding force between atoms or molecules. In order to have accurate results, AFM cantilevers must be calibrated precisely before use. The AFM cantilever's spring constant is usually provided by the manufacturer, and it is calculated from simple equations or some other calibration methods. The spring constant may have some uncertainty, which may cause large errors in force measurement. In this paper, finite element analysis was used to obtain the deformation behavior of the AFM cantilever and to calculate its spring constant. The influence of prestress, ignored by other methods, is discussed in this paper. The variations of Young's modulus, Poisson's ratio, cantilever geometries, tilt angle, and the influence of image tip mass were evaluated to find their effects on the cantilever's characteristics. The results were compared with those obtained from other methods.
Keywords:
本文献已被 PubMed 等数据库收录!