Abstract:This paper presents a comparative study of two pumping speed measurement methods recommended for oil diffusion pumps. The methods used are the test dome method and the conductance tube method with an auxiliary pumping system which is used at very low pressures when the throughput is below 10?5 Nm s?1 (~ 0.1 μl s?1). It is found that, in order to obtain the comparative results in the overlapping pressure region on the speed measurement of oil diffusion pumps by these two methods, the conductance tube in the conductance tube method should be located at a height 1.2 D from the pump mouth when the position of the inlet gauge is at a distance D/2 from the pump flange, i.e. at the same position as recommended in the dome method to obtain the intrinsic speed of the pump.