首页 | 本学科首页   官方微博 | 高级检索  
     


Tracer monitored titrations: measurement of dissolved oxygen
Authors:Martz Todd  Takeshita Yuichiro  Rolph Rebecca  Bresnahan Philip
Affiliation:University of California, San Diego, Scripps Institution of Oceanography, La Jolla, California 92093, USA. trmartz@ucsd.edu
Abstract:The tracer monitored titration (TMT) technique is evaluated for measurement of dissolved oxygen. The TMT developed in this work uses a simple apparatus consisting of a low-precision pump for titrant delivery and an optical detector based on a white LED and two photodiodes with interference filters. It is shown that the classic Winkler method can be made free of routine volumetric and gravimetric measurements by application of TMT theory, which allows tracking the amounts of titrant and sample using a chemical tracer. The measurement precision of the prototype setup was 0.3% RSD.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏