Wear tests on polypropylene for use in prosthetic heart valves
Authors:C.R. Gentle  H. Juden
Affiliation:1. Department of Mechanical Engineering, Trent Polytechnic, Burton Street, Nottingham NG1 4BU, UK
Abstract:A medical grade of polypropylene, as used in the Abrams-Lucas prosthetic mitral valve, was tested in a crossed-cylinder wear machine under conditions which reproduce those in the heart. The results show an initial dormant period of running in, followed by a steady wear rate of 2.71 × 10−5mg/m. This figure is used to give an estimate of valve life which agrees well with clinical data and confirms that the method is suitable for future use as a valve design aid without the need to resort to protracted testing of completed valves
Keywords:prosthetic devices   wear tests   polypropylene