首页 | 本学科首页   官方微博 | 高级检索  
     

用改造后的柯氏干涉仪测量量块中心长度的测量不确定度评定
引用本文:倪育才,邵宏伟,刘香斌. 用改造后的柯氏干涉仪测量量块中心长度的测量不确定度评定[J]. 现代计量测试, 1999, 7(6): 7-16
作者姓名:倪育才  邵宏伟  刘香斌
作者单位:中国计量科学研究院!北京100013
摘    要:通过用633nm 短频He-Ne 激光代替氪灯,提高小数测量的准确度等措施,对常规柯氏干涉仪进行了改造,结果获得测量量块中心长度的标准不确定度为uc= (5.5nm )2+ (8.3×10- 8L)2,大大优于一等量块的最大允许不确定度

关 键 词:柯氏干涉仪 测量不确定度 量块中心长度

Uncertainty Evaluation of the Measurement of Gauge Block by Using an Improved Koester Interferometer
NI Yu|cai,SHAO Hong|wei,LIU Xiang|bin. Uncertainty Evaluation of the Measurement of Gauge Block by Using an Improved Koester Interferometer[J]. Modern Measurement and Test, 1999, 7(6): 7-16
Authors:NI Yu|cai  SHAO Hong|wei  LIU Xiang|bin
Abstract:By the improvements such as using a stabilized He|Ne laser instead of Kr lamp,increasing the measurement uncertainty of the fraction order,etc.,the standard unceratinty of the measurement of the length of gauge block by using a Koester interferometer was obtained.It is much more better than the maximum permissible uncertainty of the 1st order gauge block.
Keywords:gauge block  Koester interferometer  measurement unceratinty  
本文献已被 CNKI 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号