In-situ high-energy-resolution X-ray absorption spectroscopy for UO2 oxidation at SSRF
Authors:Pei-Quan Duan  Hong-Liang Bao  Jiong Li  Han-Jie Cao  Yu-Ying Huang
Affiliation:Shanghai Synchrotron Radiation Facility, Shanghai Institute of Applied Physics, Chinese Academy of Sciences,Shanghai 201204, China
Abstract:Based on the high-energy-resolution fluorescence spectrometer on the BL14W1 beamline at Shanghai Synchrotron Radiation Facility,an in-situ high-energyresolution X-ray absorption spectroscopy technique,with an in-situ heating cell,was developed.The high-energyresolution fluorescence detection for X-ray absorption near-edge spectroscopy (HERFD-XANES) was tested in a UO2 oxidation experiment to measure the U L3-edge,with higher signal-to-noise ratio and higher-energy-resolution than conventional XANES.The technique has potential application for in-situ study of uranium-based materials.
Keywords:X-ray absorption spectroscopy (XAS)  In-situ cell  UO2 oxidation