Abstract:In this study, a potentiometric uric acid biosensor was fabricated by immobilization of uricase onto zinc oxide (ZnO) nanowires. Zinc oxide nanowires with 80-150 nm in diameter and 900 nm to 1.5 μm in lengths were grown on the surface of a gold coated flexible plastic substrate. Uricase was electrostatically immobilized on the surface of well aligned ZnO nanowires resulting in a sensitive, selective, stable and reproducible uric acid biosensor. The potentiometric response of the ZnO sensor vs Ag/AgCl reference electrode was found to be linear over a relatively wide logarithmic concentration range (1-650 μM) suitable for human blood serum. By applying a Nafion® membrane on the sensor the linear range could be extended to 1-1000 μM at the expense of an increased response time from 6.25 s to less than 9 s. On the other hand the membrane increased the sensor durability considerably. The sensor response was unaffected by normal concentrations of common interferents such as ascorbic acid, glucose, and urea.