The efficiency calibration of a Ge(Li) detector in short source-to-detector geometry
Wang Zhihai Fan Yangmei. The efficiency calibration of a Ge(Li) detector in short source-to-detector geometry[J]. Nuclear Techniques, 1989, 12(2): 105-109
Authors:Wang Zhihai Fan Yangmei
Affiliation:Wang Zhihai Fan Yangmei(University of Science and Technology of China)Xiao Zhenxi (Institute of Management Affiliated to Academia Sinica)
Abstract:A method for the efficiency calibration of a Ge (Li) detector at short source-to-detector distance is described. It combines the multi-gamma-ray source calibration method with the distance variation method with several single-gamma-ray sources. The full-energy peak relative efficiency, full-energy peak absolute efficiency and total efficiency have been determined with an error of 1.5%, 1.6% and 4.1% respectively at the energy ranging from 120 to 1115 keV. Therefore the spectroscopy can be used effectively to measure weak gamma-ray source which may emit cascading photons. With the method, the full-energy peak relative efficiency can be calibrated using an unknown activity ga-mma-ray source.
Keywords:Ge(Li) Efficiency calibration Short source -to-detector geometry  