首页 | 本学科首页   官方微博 | 高级检索  
     


Semiconductor and thermoluminescent dosimetry of pulsed soft X ray plasma sources
Authors:Krása J  Cejnarová A  Juha L  Ryć L  Scholz M  Kubes P
Affiliation:Department of Gas Lasers, Institute of Physics, ASCR, Prague, Czech Republic. krasa@fzu.cz
Abstract:A multichannel detection system having a dynamic range of approximately 1 x 10(-9) Gy --20 Gy was developed with the use of commercially produced Si-photodiodes and TLDs for accurate measurement of X ray energy emitted from plasma-focus facility and from laser-produced plasmas. The proof of linearity of the employed detectors accomplished by a comparison of their responses to a broad band spectrum of X rays emitted from plasmas, is reported. It is demonstrated that TLDs irradiated with no protective filter show an incorrect response due to overloading in the sub-keV range and repopulation of dosimetric peaks induced by the UV radiation. The measurement of the power of undesirable secondary X ray sources driven by the primary plasma inside the interaction chamber was performed on the basis of analysis of space dependence of X ray intensity with respect to the assumed r(-2) decrease in the intensity far away from the plasma.
Keywords: