首页 | 本学科首页   官方微博 | 高级检索  
     


Improvements in controlled-environment facilities for testing domestic air-to-water heat pumps
Authors:J. Rosell  R. Morgan  D. W. Hughes  J. T. McMullan
Abstract:A control system is described, based on a microcomputer, which maintains constant temperature and relative humidity in a controlled-environment laboratory used for research on air-to-water heat pumps. The performance of the control system is measured under both steady-state and changing conditions, and optimum values of the control constants are deduced. The temperature can be maintained at the set value to within 0–1°C, and the relative humidity to within 1 per cent, under steady-state conditions. The effects of transients, such as the starting or stopping of the heat pump under test, or the deliberate changing of the controlled conditions, are studied in detail. In addition, a system is described for controlling the temperature of the condenser cooling water, which can be held to within 0–1°C of the set value.
Keywords:Heat pump  Test facilities  Controlled-environment laboratory
设为首页 | 免责声明 | 关于勤云 | 加入收藏