Helium‐Testlecks |
| |
Authors: | Dr Wolfgang Jitschin Dr Baker Farangis |
| |
Affiliation: | Labor für Vakuumtechnik, Fachbereich MNI, TH Mittelhessen, Wiesenstr. 14, 35390 Gie?en |
| |
Abstract: | Helium test leaks Leak localization and tightness control with Helium provides a very sensitive testing method. Frequently, the sensitivity of a leak has to be testspecified quantitatively. In particular, quality control requires a reliable verification of the sensitivity. For this purpose, test leaks with well‐known leakage are used. In the article, various types of test leaks and their properties are described. Test leaks are commercially available over a wide range of leak rates. Their leakage flow can be calibrated with good accuracy. In the usage of the test leaks, the temperature‐dependency of the leak rate has to be taken into account, as well as the slow decrease of the leak rate due to the continuous loss of gas from the reservoir. |
| |