首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement of acetaldehyde in exhaled breath using a laser absorption spectrometer
Authors:Kamat Pratyuma C  Roller Chad B  Namjou Khosrow  Jeffers James D  Faramarzalian Ali  Salas Rodolfo  McCann Patrick J
Affiliation:Ekips Technologies, Incorporated, Norman, Oklahoma 73069, USA.
Abstract:A high-resolution liquid-nitrogen-free mid-infrared tunable diode laser absorption spectroscopy (TDLAS) system was used to perform real-time measurement of acetaldehyde concentrations in human exhaled breath following ingestion of an alcoholic beverage. Acetaldehyde absorption features were measured near 5.79 mum (1727 cm(-1)) using a IV-VI semiconductor laser, a 100 m long path optical gas cell, and second- harmonic detection coupled with wavelength modulation. Acetaldehyde levels were measured with a minimum detection limit of 80 ppb for 5 s integration time. The variations in exhaled acetaldehyde levels over time were analyzed prior to and following ingestion of two different amounts of white wine. A method to calibrate acetaldehyde measurements internally using water vapor absorption lines was investigated to eliminate the need for system calibration with gas standards. The potential of a TDLAS system to be used as a noninvasive clinical tool for measurements of large volatile compounds with possible applications in cancer detection is demonstrated.