首页 | 本学科首页   官方微博 | 高级检索  
     


Pulse oximetry theory and calibration for low saturations
Authors:Zonios George  Shankar Uday  Iyer Vijay K
Affiliation:Respironics, Inc., Kennesaw, GA 30144, USA.
Abstract:Pulse oximetry is a widely used technique in biomedical optics, but currently available pulse oximeters rely on empirical calibration approaches, which perform poorly at low saturations. We present an exact solution for pulse oximetry and show how this can be used as the basis for the development of a semiempirical calibration approach that may be useful, especially at low saturations and variable probe geometries. This new approach was experimentally tested against traditional empirical calibration techniques on transmission pulse oximetry for monitoring of fetal sheep using a minimally invasive spiral probe. The results open the way for the development of more accurate pulse oximetry.
Keywords:
本文献已被 PubMed 等数据库收录!