首页 | 本学科首页   官方微博 | 高级检索  
     

加权平均法表述菌落总数测定的数学模型与 测量不确定度的评定
引用本文:易大为,高敬红,邹 宇. 加权平均法表述菌落总数测定的数学模型与 测量不确定度的评定[J]. 食品安全质量检测学报, 2014, 5(9): 2888-2892
作者姓名:易大为  高敬红  邹 宇
作者单位:辽宁省药品检验检测院,沈阳志鹰药业有限公司,辽宁省药品检验检测院
摘    要:摘 要 目的:建立食品菌落总数测定的通用数学模型和测量不确定度的评定方法。方法:按照加权平均的算法对菌落总数计算公式进行推导,建立了菌落总数计数方法的通用数学模型。以某橙汁饮料菌落总数的测定过程为例,从三个方面分析并量化了各不确度的分量,分别为:不同稀释度平板上菌落数的分布(以同一稀释度上两个平板计数结果的相对相差表示),样品移取和稀释过程(以各量具的容量允差及稀释过程对稀释倍数贡献的相对不确度表示),数字修约(以修约间隔引入的粗大误差表示)。按照均方根法则,将各计数平板的测量不确度进行加和,即为总测量不确度。结果:本例中该橙汁饮料菌落总数的扩展不确度U=0.2×104 CFU/mL, k =2。结论:建立了简便准确的食品菌落数测量不确定度的评定方法。

关 键 词:不确定度;菌落总数;加权平均
收稿时间:2014-08-12
修稿时间:2014-09-09

Establishment of mathematical model and evaluation of measurement uncertainty for Aerobic Plate Count by weighted average
YI Da-Wei,GAO Jing-Hong and ZOU Yu. Establishment of mathematical model and evaluation of measurement uncertainty for Aerobic Plate Count by weighted average[J]. Journal of Food Safety & Quality, 2014, 5(9): 2888-2892
Authors:YI Da-Wei  GAO Jing-Hong  ZOU Yu
Affiliation:Liaoning Institute for Drug Control,Shenyang Zhiying Pharmaceutical Co Ltd,Liaoning Institute for Drug Control
Abstract:Objective To establish the general mathematical model and the procedure of measurement uncertainty for the determination of the Aerobic Plate Count (APC) on food. Methods According to the weighted average method, the formula of the APC was derived, and a general mathematical model was established. To take the measurement process of an orange juice drink APC as an example, three uncertainty com-ponents were evaluated and expressed, including the distribution of colony forming units (CFUs) on selected plates at different dilutions (equal to the relative difference of CFUs at the same dilution plates), the sampling and diluting process (equal to the relative uncertainty of dilution factor from capacity allowable errors and dilution process cumulative errors), rounding off numerical values (equal to gross error of rounding interval). According to the root-mean-square principle, measurement uncertainty components of each plate were added to calculate the total measurement uncertainty. Results Expanded uncertainty for the APC is 0.2×104 CFU/mL, k=2. Conclusion A simple and accurate method for evaluating measurement uncertainty for the APC of food is established.
Keywords:uncertainty   aerobic plate count   weighted average