首页 | 本学科首页   官方微博 | 高级检索  
     

取样示波器nose-to-nose校准技术中kick-out脉冲的研究
引用本文:朱江淼,刘明亮,高昀.取样示波器nose-to-nose校准技术中kick-out脉冲的研究[J].中国计量学院学报,2003,14(3):185-188.
作者姓名:朱江淼  刘明亮  高昀
作者单位:北京工业大学,电子信息与控制工程学院,北京,100022
摘    要:研究了nose-to-nose校准技术中关键信号kick-out脉冲产生的机理以及影响因素,并通过将两台Agilent86100系列的50GHz的取样示波器对接,获得了kick-out脉冲的响应波形.

关 键 词:取样示波器  kick-out脉冲  冲激响应  nose-to-nose校准
文章编号:1004-1540(2003)03-0185-04

The study of the kick-out Pulse in sampling oscilloscope's nose-to-nose calibration procedure
ZHU Jiang\|miao,LIU Ming\|liang,GAO Yun.The study of the kick-out Pulse in sampling oscilloscope''s nose-to-nose calibration procedure[J].Journal of China Jiliang University,2003,14(3):185-188.
Authors:ZHU Jiang\|miao  LIU Ming\|liang  GAO Yun
Abstract:Kick\|out pulse is the key signal in the nose\|to\|nose calibration procedure, the premise of the NTN technology is that kick\|out pulse is generated when sampling head samples a DC voltage and the kick\|out pulse is proportion to the impulse response of the sampling oscilloscope, mechanism of the kick\|out pulse generation and related effects on it have been studied in this paper. By using two Agilent 86100 sampling oscilloscopes nose\|to\|nose, we obtain the response waveform of the kick\|out pulse.
Keywords:sampling oscilloscope  kick-out pulse  impulse response  nose-to-nose calibration