首页 | 本学科首页   官方微博 | 高级检索  
     


A theoretical and photoelastic contact stress analysis relevant to orthopaedic knee prostheses
Authors:M. P. Whelan  E. G. Little
Affiliation:Department of Mechanical and Production Engineering, University of Limerick, Plassey Technological Park, Limerick, treland.
Abstract:This study is based on a theoretical and photoelastic static contact stress analysis, relevant to the design of the plastic tibial plateau in an orthopaedic knee prosthesis. It involves the review of existing theoretical analyses concerned with the contact mechanics of layered elastic systems and, where possible, the validation of any relevant findings using two dimensional photoelasticity. Significant findings include the identification of a general minimum recommendable design thickness for the plastic tibial plateau, the endorsement of using the Hertzian contact theory to predict contact stress levels in the tibial plateau of certain knee implants and the slight improvement in contact conditions in the plastic when the plateau is perfectly bonded to its supporting metal tray.
Keywords:Contact    stress    theoretical    photoelastic    prostheses.