首页 | 本学科首页   官方微博 | 高级检索  
     


Mean systemic pressure and mean pulmonary pressure: their effects on the total artificial heart
Authors:Y Kito  T Honda  WH Gibson  T Akutsu
Abstract:Mean systemic pressure (MSP) and mean pulmonary pressure (MPF), which are mean driving pressures for venous return in the natural heart, were studied in 11 calves in which the natural heart had been replaced with a total artificial heart (TAH). They were measured simply by stopping the artificial heart pumping. Although blood translocation from the arterial to the venous side was not performed, the eventual right and left atrial pressures reached six to eight seconds after stopping the TAH would represent MSP and MPP with reasonable accuracy. The MSP varied from nine to 3k mmHg (20+/-6 mmHg), whereas the MPP varied from nine to 39 mmHg (22+/-7 mmHg). The MSP varied in close relation to the right atrial pressure prior to cessation of the TAH (r=0.9124). Increases in RAP and MSP were mainly attributed to an increase in circulating blood volume. In the performance of the TAH, MSP (or MPP), proper diastolic duration and vacuum application during diastole was of prime importance in determining the end-diastolic ventricular volume.