首页 | 本学科首页   官方微博 | 高级检索  
     


Characterization of Two Ti-Nb-Hf-Zr Alloys Under Different Cold Rolling Conditions
Authors:M Gonz??lez  F J Gil  J M Manero  J Pe?a
Affiliation:1. Department of Materials Science and Metallurgy, Universitat Polit??cnica de Catalunya, Avda. Diagonal 647, 08028, Barcelona, Spain
2. Materials Science, Elisava Escola Superior de Disseny, C/Ample 11-13, 08002, Barcelona, Spain
Abstract:The objective of this study is to obtain a new biomaterial to use in the load transfer medical implants field. The influence of cold work in the thermoelastic martensitic transformation and elastic modulus was investigated for a previously developed Ti-24.8Nb-16.2Hf-1Zr alloy (A1) and a new Ti-35Nb-9Hf-1Zr alloy (A2). The nanoindentation tests with spherical tip showed a decrease of the elastic modulus when increasing the cold work percentage. The lowest value (46 GPa) was achieved after the 95% reduction in thickness for A1. By contrast, the A2 alloy exhibits its lowest elastic modulus in the uncold-worked condition (58 GPa) and did not present evidences of thermoelastic martensitic transformation.