首页 | 本学科首页   官方微博 | 高级检索  
     


An Investigation of the Linearity of Standard Thermistor Wattmeters and an Estimate of the Minimum Attainable Errors of Measurements of the Ratio of Microwave Powers
Authors:Perepelkin  V A  Pivak  A V  Chuiko  V G
Abstract:The results of experimental investigations of the nonlinearity of thermistor power converters are presented and the main components of the error in measuring nonlinearity in a differential microcalorimeter are analyzed. The resultant error in determining nonlinearity does not exceed 0.002 dB/10 dB at frequencies up to 18 GHz.
Keywords:microwave power  thermistor  linearity  attenuation  microcalorimeter
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏