首页 | 本学科首页   官方微博 | 高级检索  
     


The calibration temperature of pH-glass electrodes: significance for meat quality classification
Authors:Karlsson Anders H  Rosenvold Katja
Affiliation:

Department of Animal Product Quality, Danish Institute of Agricultural Sciences, PO Box 50, DK-8830, Tjele, Denmark

Abstract:In this short communication we show important factors concerning pH measurements in pork. The purpose of the study was to investigate the effect of the calibration temperature of pH-buffer and of the pH-electrode on pH measured 1–120 min post mortem in pork, and to evaluate its significance for prediction of meat quality. The calibration temperature of pH-electrodes as well as the sample temperature is essential for pH measurements, as pH measurements are highly temperature-dependent. This is especially important when using pH measurement for prediction of meat quality or when different experiments are compared. We have used two calibration temperatures of 15 °C (normal) and 35 °C (modified) in the present investigation. Based on our findings, we therefore recommend a buffer and electrode temperature of 35 °C as a standard procedure for measuring pH early post mortem. As a minimum the pH should be presented along with the measuring and calibrating temperatures.
Keywords:Calibration   Classification   Meat quality   Pig   pH
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏