首页 | 本学科首页   官方微博 | 高级检索  
     

MAT261型离子源及离子透镜系统的仿真与实际应用
引用本文:韦冠一,翟利华,张子斌,王长海,李雪松,万可友.MAT261型离子源及离子透镜系统的仿真与实际应用[J].质谱学报,2005,26(Z1):99-100.
作者姓名:韦冠一  翟利华  张子斌  王长海  李雪松  万可友
作者单位:西北核技术研究所,陕西,西安,710024
摘    要:

文章编号:1004-2997(2005)增刊-99-02
修稿时间:2005年7月10日

Simulation of MAT-261 Type Ion Source and Its Application
WEI Guan-yi,ZHAI Li-hua,ZHANG Zi-bin,WANG Chang-hai,LI Xue-song,WAN Ke-you.Simulation of MAT-261 Type Ion Source and Its Application[J].Journal of Chinese Mass Spectrometry Society,2005,26(Z1):99-100.
Authors:WEI Guan-yi  ZHAI Li-hua  ZHANG Zi-bin  WANG Chang-hai  LI Xue-song  WAN Ke-you
Abstract:The simulation results of MAT-261 type ion source and ion optics with the SIMION 3D 7.0 software were given. By means of simulation, the ion's trajectory was studied, the cause of low abundance sensitivity was analyzed and the optimal poles' voltage ranges of the ion optics were found. Modifications were made to the LRIMS3000 type mass spectrometer according to result of the simulation. The abundance sensitivity of the mass spectrometer is improved to 2×10-6 from original 5×10-5.
Keywords:ion-optics  simulation  abundance sensitivity  mass spectrometer
本文献已被 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号