首页 | 本学科首页   官方微博 | 高级检索  
     


Effect of Methylal Quenching Gas for Self-Quenching Streamer (SQS) Tube with Ar-isoC4H10-Methylal Mixture
Abstract:Abstract

Properties of a self-quenching streamer (SQS) tube with an Ar-isoC4H10 (75:25) mixture containing 2.5% methylal vapor have been investigated for uniform α-ray irradiation. The effect of electro-negative methylal as the second quenching gas has been studied for avalanche sizes, pulse shapes of current signals, dead zones, and counting-rate characteristics of the SQS tube. The dead zone and counting-rate characteristics are improved very much by the use of the mixture containing methylal.
Keywords:gas counter  self-quenching streamer mode  self-quenching streamer tube  GM tube  gamma radiation  radiation monitors  methylal  counting rates