首页 | 本学科首页   官方微博 | 高级检索  
     

火焰原子吸收分光光度法测定铜操作条件的选择
引用本文:盛丽,苏碧泉. 火焰原子吸收分光光度法测定铜操作条件的选择[J]. 辽宁化工, 2006, 35(9): 557-559
作者姓名:盛丽  苏碧泉
作者单位:兰州交通大学化工学院,甘肃,兰州,730070;兰州交通大学化工学院,甘肃,兰州,730070
摘    要:在单因素试验的基础上,采用四因素四水平正交试验设计测试WFX-120原子吸收分光光度计测定铜的操作参数。依据L16(45)正交表,得出最佳工作条件为灯电流1.6 mA、乙炔流量/空气流量1.0/6.5 L/min、狭缝宽度0.4 mm、燃烧器高度5 mm。

关 键 词:原子吸收分光光度法    正交试验设计
文章编号:1004-0935(2006)09-0557-03
收稿时间:2006-05-09
修稿时间:2006-05-09

The Selection of Operating Conditions of Determinating the Copper Ion by Flame Atomic Absorption
SHENG Li,SU Bi-quan. The Selection of Operating Conditions of Determinating the Copper Ion by Flame Atomic Absorption[J]. Liaoning Chemical Industry, 2006, 35(9): 557-559
Authors:SHENG Li  SU Bi-quan
Affiliation:School, of Chemical and Biological Engineering, Lanzhou Jiaotong University, Lanzhou 730070, China
Abstract:On the base of the single factor experiment,optimum operation conditions of determinating the copper ion by flame atomic absorption were researched by means of orthogonal test with four factors and four levels.In the orthogonal layout of L_(16)(4~5),the results show that optimum operation conditions of determinating the copper ion by flame atomic absorption on WFX-120 atomic absorption spectrometer are that lamp current is 1.6mA,width of slit is 0.4mm,height of the burnet 5mm and the ratio of flowing of acetylene to flow of air 1.0/6.5 L/min.
Keywords:Flame atomic absorption spectrophotometry  Copper ion  Orthogonal test design  
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏