首页 | 本学科首页   官方微博 | 高级检索  
     

DETERMINATION OF WHOLE—BODY PROTEIN IN SMALL LIVE ANIMALS BY IVNAA
作者姓名:王海婴  Charles M. Waana
作者单位:Department of Physics University of Toronto Canada,Department of Physics University of Toronto Canacla
摘    要:Measurement of nitrogen in small live animals (Weighing about 3 kg) can be used to determine the whole- body protein in the animals. Nitrogen may be measured by IVNAA in which the capture gamma rays of neutron induced are detected and counted. The neutron flux is provided by a collimated 740 GBq (20Ci) Pu- Be source. The 10.83 MeV thermal neutron capture gamma rays from ~(14)N are detected by a 12.7 cm by 10.2 NaI (T1) detector. The nitrogen of a live rabbit was measured and recorded each day for a period of two months. The statistical error and reproducibility of measurement were around 10%. For a 1000 s irradiation the dose equivalent was 300μ Sv.

本文献已被 CNKI 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏