首页 | 本学科首页   官方微博 | 高级检索  
     


Error grid analysis of noninvasive glucose monitoring via gingival cerevicular fluid
Authors:Yamaguchi  M Kambe  S Yamazaki  K Kobayashi  M
Affiliation:Fac. of Eng., Toyama Univ., Japan;
Abstract:In order to realize a noninvasive blood glucose monitor, we monitored the gingival cerevicular fluid (GCF). In this paper, the clinical evaluations were performed on not only normal subjects but also diabetic subjects using a GCF-glucose monitor to determine blood glucose levels. Meal load tests were carried out and the time-course changes in blood glucose level and GCF glucose level were measured continuously. A positive correlation of more than 0.9 was found between blood glucose level and GCF glucose level, necessitating the calibration of individual correlations for every subject. Finally, the performance of the GCF-glucose monitor was evaluated using Error Grid Analysis. As the results, significant information for the glucose level decision was obtained not only for normal subjects, but also for diabetic subjects.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号