Heat transfer evaluation of the nasal thermistor technique |
| |
Authors: | K Storck M Karlsson P Ask D Loyd |
| |
Affiliation: | Department of Mechanical Engineering Link?ping University, Sweden. |
| |
Abstract: | When analyzing transvalvular and venous flow velocity patterns, it is important to relate them to respiration. An accurate recording of respiratory phase can be carried out with different methods. One of these methods is the use of a thermistor, which reacts to the variation in air temperature, placed in the noise of the patient. The thermistor used has a diameter for 1.0 mm and is of standard bead type. Although small, it has a considerable long time-constant and a long time-delay. The high time-constant gives a low cutoff frequency, well below the respiratory frequency and thereby causing a large phase difference. The thermistor was analyzed with the lumped heat capacity method, where it was easy to study the influence from design parameters, time-dependent air temperature, and velocity. The analysis was extended using the finite element method and the temperature field in the thermistor and the probe was calculated as a function of space and time. These calculations confirmed the result from the lumped model. The result showed that timing of respiration was not accurately obtained with the thermistor analyzed. To improve the timing, it was necessary either to change the measuring method or to use signal processing in order to achieve faster response. |
| |