首页 | 本学科首页   官方微博 | 高级检索  
     

X射线环境辐射监测仪器校准方法的研究
引用本文:宋飞,赵瑞,丁卫撑,张德亮,杨扬,吴金杰. X射线环境辐射监测仪器校准方法的研究[J]. 计量学报, 2021, 42(9): 1237-1243. DOI: 10.3969/j.issn.1000-1158.2021.09.19
作者姓名:宋飞  赵瑞  丁卫撑  张德亮  杨扬  吴金杰
作者单位:成都理工大学核技术与自动化工程学院,四川成都610059;中国计量科学研究院,北京100029;中国计量科学研究院,北京100029;成都理工大学核技术与自动化工程学院,四川成都610059
基金项目:国家重点研发计划(2017YFF0205101)
摘    要:环境辐射监测仪表作为微弱放射性监测计量器具,它的量值溯源是一个急需解决的问题.由于环境辐射剂量率低且电离信号微弱,故通常采用大体积电离室或者充压电离室进行测量.依托60 ~250 kV X射线空气比释动能基准电离室,在完成重过滤窄谱X射线辐射质空气比释动能绝对测量的基础上,通过逐级替代法完成大体积环境辐射监测仪器的校准...

关 键 词:计量学  X射线  空气比释动能  逐级替代法  溯源
收稿时间:2021-02-22

Research on Calibration Method of X-ray Environmental Radiation Monitoring Instrument
SONG Fei,ZHAO Rui,DING Wei-cheng,ZHANG De-liang,YANG Yang,WU Jin-jie. Research on Calibration Method of X-ray Environmental Radiation Monitoring Instrument[J]. Acta Metrologica Sinica, 2021, 42(9): 1237-1243. DOI: 10.3969/j.issn.1000-1158.2021.09.19
Authors:SONG Fei  ZHAO Rui  DING Wei-cheng  ZHANG De-liang  YANG Yang  WU Jin-jie
Affiliation:1. College of Nuclear Technology and Automation Engineering,Chengdu University of Technology, Chengdu, Sichuan 610059, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:As a measuring instrument for weak radioactivity monitoring, the traceability of the environmental monitoring is an urgent problem to be solved. Because the environmental radiation dose rate is low and the ionization signal is weak, it is usually measured by large volume ionization chamber or charging pressure ionization chamber. On the based of primary standards of 60~250kV X-ray air kerma and absolute measurement of the air kerma under the narrow spectrum series X-ray radiation quality, radiation environmental monitoring instruments were calibratied by successive substitution method. The uncertainty of measurement measurement result is better than 5.6%(k=2). It realize the traceability of environmental level X-ray air kerma measurement, and provide metrological support for the performance evaluation of environmental radiation monitoring instruments at low dose rate level.
Keywords:metrology  X-ray  air kerma standard  successive substitution method  traceability  
本文献已被 万方数据 等数据库收录!
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号