基于光路折叠的室内基线场误差分析方法研究
引用本文:蔡青梅,赵美蓉,刘红光,郑叶龙,张涵莅,宋乐.基于光路折叠的室内基线场误差分析方法研究[J].计量学报,2021,42(9):1142-1148.
作者姓名:蔡青梅  赵美蓉  刘红光  郑叶龙  张涵莅  宋乐
作者单位:天津大学精密仪器与光电子工程学院,天津300072;天津市计量监督检测科学研究院,天津300192
基金项目:国家自然科学基金(51805367);天津市自然科学基金(18JCQNJC04800, 17JCYBJC19000和18JCZDJC31800)
摘    要:为了有效开展长距离光电测距类仪器的室内检定工作,在50m高精度导轨上利用1组平面反射镜进行光路折叠,搭建了100m室内基线场。对基线系统的测量误差进行了分析,考虑各项误差的调整精度,将误差分析结果应用到实际光路调整过程中,分析得到对基线长度影响更显著的误差量,并对其进行了控制和调整,提高测量光路的平行性调整精度,最后选择双频激光干涉仪作为长度基准开展验证实验。实验结果表明:通过光路折叠方法可以实现2倍光程倍增,基线系统的精度较高,可进行连续测量,有效地解决了室内基线建立过程中所存在的检测效率低、可重复性差等技术问题。

关 键 词:计量学  大尺寸测量  光路折叠  室内基线场  双频激光干涉仪  平面反射镜  误差分析
收稿时间:2020-06-12

Error Analysis Methodology for Indoor Baseline Field Based on Optical Path Folding
CAI Qing-mei,ZHAO Mei-rong,LIU Hong-guang,ZHENG Ye-long,ZHANG Han-li,SONG Le.Error Analysis Methodology for Indoor Baseline Field Based on Optical Path Folding[J].Acta Metrologica Sinica,2021,42(9):1142-1148.
Authors:CAI Qing-mei  ZHAO Mei-rong  LIU Hong-guang  ZHENG Ye-long  ZHANG Han-li  SONG Le
Affiliation:1. Precision Instruments and Optoelectronic Engineering College, Tianjin University, Tianjin 300072, China
2. Tianjin Institute of Metrological Supervision and Testing, Tianjin 300192, China
Abstract:In order to effectively carry out indoor verification of long-distance photoelectric distance measuring instruments, an indoor virtual baseline field was built based on the principle of folding optical path with a group of plane mirrors. The measurement error of the baseline system was analyzed. Considering the adjustment accuracy of various errors, the analysis results were applied to the actual optical path adjustment process. Through the analysis, the errors which have more significant effect on the baseline length were obtained, and they were controlled and adjusted to improve the adjustment accuracy of the parallelism of the measuring optical path. Finally, a dual frequency laser interferometer was used to provide reference length value to carry out the verification experiment. The experimental results show that double optical path multiplication can be achieved by optical path folding method, and the baseline system has high accuracy and can be used for continuous measurement, effectively solving the technical problems of low detection efficiency and poor repeatability in the process of indoor baseline establishment.
Keywords:metrology  large size measurement  optical path folding  indoor baseline field  dual frequency laser interferometer  plane mirror  error analysis  