APD探测器的性能对激光雷达反演CO2浓度误差影响研究
引用本文:左维康,朱亚丹,邱敏,刘继桥,陈卫标.APD探测器的性能对激光雷达反演CO2浓度误差影响研究[J].红外与激光工程,2018,47(4):406002-0406002(7).
作者姓名:左维康  朱亚丹  邱敏  刘继桥  陈卫标
作者单位:1.中国科学院上海光学精密机械研究所 中科院空间激光信息传输与探测技术重点实验室,上海 201800;
基金项目:民用航天项目(1105231-CAO)
摘    要:星载积分路径差分吸收(IPDA)激光雷达是全天时全球范围内探测CO2浓度的一种有效的方法,而作为接收系统关键元件的光电探测器对激光雷达系统性能有着较大的影响。雪崩光电二极管(APD)有着较大的动态范围与高的响应度,因此它在星载激光雷达中广泛应用。介绍了IPDA激光雷达和APD探测器的工作原理,并根据实际工作条件,测试了一款APD探测单元的响应度、动态范围、不同光功率下的信噪比等主要性能参数,分析了这些性能参数对星载激光雷达CO2浓度的反演带来的影响。结果表明,在CO2浓度为400 ppm(1 ppm=10-6),吸收波段信号的探测器输出电压在280~980 mV范围内时,APD探测器本身的非线性和噪声造成的误差小于0.8 ppm。

关 键 词:误差分析    积分路径差分吸收激光雷达    APD探测器    非线性    信噪比
收稿时间:2017-11-08

Influence of performances of APD detector on CO2 concentration error retrieved by lidar measurement
Affiliation:1.Key Laboratory of Space Laser Communication and Detection Technology,Shanghai Institute of Optics and Fine Mechanics,Chinese Academy of Science,Shanghai 201800,China;2.University of Chinese Academy of Sciences,Beijing 100049,China
Abstract:Space-borne integrated path differential absorption(IPDA) lidar has been regarded as an efficient method to measure the concentration of the global atmospheric CO2 column concentration all day long. As a key part of the lidar receiver, the performance of the photodetector has a significant influence in the overall lidar system. Avalanche photodiode(APD) is widely used in space-borne lidar because of its large dynamic range and high responsivity. The basic theories of the APD detector and the IPDA lidar were briefly introduced. And taking the actual working conditions of the space-borne IPDA lidar into account, the actual performances of an APD detector such as the detector's responsivity, dynamic range, linearity and the signal to noise ratio under different optical powers were tested. Then based on the experiment result, the influence caused by the performances of the detector on concentration inversion was analyzed. The result shows that, when the CO2 concentration is 400 ppm, and the on-line wavelength output voltage of the APD detector is between 280 mV and 980 mV, the CO2 concentration error caused by the non-linearity and noise of the APD detector is less than 0.8 ppm.
Keywords: