基于光纤腔衰荡光谱的浓度和温度测量 |
| |
引用本文: | 程瑞学,芦恒,杨亚萍,王芳. 基于光纤腔衰荡光谱的浓度和温度测量[J]. 激光技术, 2018, 42(4): 482-486. DOI: 10.7510/jgjs.issn.1001-3806.2018.04.010 |
| |
作者姓名: | 程瑞学 芦恒 杨亚萍 王芳 |
| |
作者单位: | 1.新乡市中心医院 影像中心, 新乡 453000 |
| |
摘 要: | 为了提高浓度和温度测量的灵敏度和稳定性,采用时域分析法监测光纤系统中的光损耗,研制了基于光纤环形腔衰荡光谱的传感系统。基于该系统对浓度和温度进行传感测量实验,分析了错位传感结构的参量选择,并研究了空载时腔内信号放大对脉冲强度和脉冲数量的影响。结果表明,当干涉长度L和错位量D分别为4cm和3.75μm时,干涉效果最优;脉冲强度是无腔内放大时的4倍且脉冲数量更多;当蔗糖和葡萄糖溶液浓度为0.100g/mL~0.400g/mL时,浓度灵敏度为756.51μs/(g/mL)和909.07μs/(g/mL),检测限为0.0014g/mL;当温度为30℃~200℃时,温度灵敏度为1.83μs/℃。该系统的设计和研究为浓度和温度的传感应用提供了有价值的指导。
|
关 键 词: | 传感器技术 光纤环形腔衰荡 干涉 芯偏移 |
收稿时间: | 2017-09-22 |
Measurement of concentration and temperature based on fiber loop ring-down spectroscopy |
| |
Abstract: | In order to improve the sensitivity and stability of concentration and temperature measurement, the optical loss in the optical fiber system was monitored by means of time-domain analysis. A sensing system based on fiber loop ring-down spectroscopy (FLRDS) was proposed. Based on the measurement results of concentration and temperature, parameter selection of the core-offset sensing structure was analyzed experimentally. The effect of intra-cavity signal amplification on pulse intensity and pulse number was studied. The results show that, when the interference length L and offset D are 4cm and 3.75μm respectively, interference effect is optimal. Pulse peak intensity is 4 times that without intra-cavity amplification and pulse number is bigger. When the sucrose and glucose concentration is 0.100g/mL~0.400g/mL, concentration sensitivities are 756.51μs/(g/mL) and 909.07μs/(g/mL), and the detection limit is 0.0014g/mL. The temperature sensitivity is 1.83μs/℃ in the range of 30℃~200℃. The design and research of this system provide valuable guidance for sensing application of concentration and temperature. |
| |
Keywords: | |
|
| 点击此处可从《激光技术》浏览原始摘要信息 |
|
点击此处可从《激光技术》下载免费的PDF全文 |
|