首页 | 本学科首页   官方微博 | 高级检索  
     

小型化燃煤电厂逃逸氨气激光测量仪
引用本文:邱选兵,魏永卜,孙冬远,张恩华,李传亮,魏计林,臧振中.小型化燃煤电厂逃逸氨气激光测量仪[J].激光技术,2019,43(5):697-701.
作者姓名:邱选兵  魏永卜  孙冬远  张恩华  李传亮  魏计林  臧振中
作者单位:太原科技大学 应用科学学院,太原 030024;江西中医药大学 现代中药制剂教育部重点实验室,南昌 330004;太原科技大学 应用科学学院,太原,030024;江西中医药大学 现代中药制剂教育部重点实验室,南昌,330004
基金项目:国家自然科学基金;国家自然科学基金;太原科技大学博士启动基金;教育部重点实验室开放基金;教学改革研究项目
摘    要:为了准确测量高温下逃逸氨的体积分数,采用可调谐半导体激光吸收光谱技术、波长调制光谱技术(WMS)和长光程技术,开发了一套高温小型化的逃逸氨气测量仪;为了减小逃逸氨气的吸附效应和提高探测灵敏度,研制了新型高温长光程样品吸收池。在前期研制的激光驱动模块基础上,采用74HC4046锁相环芯片作为可调正弦调制信号源,以EPM7064为移相和倍频逻辑控制芯片,同时采用两片AD630作为一次解调(WMS-1f)和二次解调(WMS-2 f)同步解调乘法器,实现了吸收信号的1 f和2f同步解调。此外,以STM32F429为主控制器,将解调滤波后的信号输入到AD7606进行模数转换,并进行数字滤波和体积分数的反演。结果表明,氨气体积分数与WMS-1f幅值、WMS-2f幅值以及同步解调WMS-2 f/WMS-1 f归一化幅度值的线性拟合系数分别为0.998,0.997以及0.998;Allen方差表明在优化时间228s时,其测量的体积分数最低为0.496×10-6,在体积分数为20×10-6~100×10-6范围内测量误差小于±2%。该测量仪可以为燃煤电厂氨逃逸的高温测量提供高精度的原始数据。

关 键 词:光谱学  氨气激光测量仪  可调谐激光光谱  模拟锁向放大器  波长调制  高温多光程池
收稿时间:2018-10-11

A miniaturized laser measurement instrument of ammonia escaping from coal-fired power plants
QIU Xuanbing,WEI Yongbo,SUN Dongyuan,ZHANG Enhua,LI Chuanliang,WEI Jilin,ZANG Zhenzhong.A miniaturized laser measurement instrument of ammonia escaping from coal-fired power plants[J].Laser Technology,2019,43(5):697-701.
Authors:QIU Xuanbing  WEI Yongbo  SUN Dongyuan  ZHANG Enhua  LI Chuanliang  WEI Jilin  ZANG Zhenzhong
Affiliation:(School of Applied Science,Taiyuan University of Science and Technology,Taiyuan 030024,China;Key Laboratory of Modern Preparation of Traditional Chinese Medicine,Ministry of Education,Jiangxi University of Traditional Chinese Medicine,Nanchang 330004,China)
Abstract:In order to accurately measure volume fraction of the escaping ammonia at high temperature, tunable diode laser absorption spectroscopy, wavelength modulation spectroscopy (WMS) and long-path technology were used. A set of miniaturized measurement instrument for escaped ammonia gas at high temperature was developed. In order to reduce the adsorption effect of escaping ammonia and improve the detection sensitivity, a new type of sample absorption cell with high temperature and long optical path was developed. On the basis of the laser driving module developed earlier, phase-locked loop chip 74HC4046 was used as an adjustable sinusoidal modulation signal source. EPM7064 was used as phase shifting and frequency doubling logic control chip. At the same time, two AD630 chips were used as simultaneous demodulation multiplier of the first demodulation (WMS-1f) and the second demodulation (WMS-2f). The synchronous demodulation of 1f and 2f absorption signals was realized. In addition, STM32F429 was used as main controller and the demodulated and filtered signal was input to AD7606 for analog-to-digital conversion. Digital filtering and volume fraction inversion were performed. The results show that linear fitting coefficients of volume fraction of ammonia gas with WMS-1f amplitude, WMS-2f amplitude and the normalized amplitude of WMS-2f/WMS-1f are 0.998, 0.997 and 0.998, respectively. Allen variance shows that when the optimization time is 228s, the lowest volume fraction is 0.496×10-6. In the range of volume fraction 20×10-6~100×10-6, the measurement error is less than ±2%. The instrument can provide high precision raw data for high temperature measurement of ammonia escaping from coal-fired power plants.
Keywords:spectroscopy  ammonia laser measurement instrument  tunable laser spectroscopy  analog lock-in amplifier  wavelength modulation  high temperature multi-path cell
本文献已被 维普 万方数据 等数据库收录!
点击此处可从《激光技术》浏览原始摘要信息
点击此处可从《激光技术》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号