首页 | 本学科首页   官方微博 | 高级检索  
     


A method of RF inhomogeneity correction in MR imaging
Authors:Hiroaki Mihara  Norio Iriguchi  Shogo Ueno
Affiliation:(1) Graduate School of Medicine, University of Tokyo, 7-3-1 Hongou, Bunkyouku, 113 Tokyo, Japan
Abstract:A direct postprocessing method for correcting RF inhomogeneity in MR imaging is proposed. First, two images with different flip-angles of ψ and 2ψ are obtained. Next, the spatial distribution maps of the sensitivity of the surface coil and theB 1 field intensity are produced by employing those images. Finally, the correction of the MR image is achieved, dividing the original image by distribution maps of the coil sensitivity and theB 1 field intensity. The method was applied to image obtained by a gradient echo sequence and the corrected image is presented.
Keywords:MRI  RF inhomogeneity correction  Post-acquisition inhomogeneity correction  Flip angle  Surface coil