首页 | 本学科首页   官方微博 | 高级检索  
     

电感耦合等离子体质谱法( ICP-MS)测定地下水中痕量元素
引用本文:蔡述伟,程青.电感耦合等离子体质谱法( ICP-MS)测定地下水中痕量元素[J].净水技术,2011(4):56-58.
作者姓名:蔡述伟  程青
作者单位:南京地质矿产研究所;
摘    要:用电感耦合等离子体质谱法(ICP-MS)测定了地下水中痕量元素(Cd、Co、Cr、Cu、Mn、Mo、Ni、Pb、As、Se、Zn、Ba、Fe),应用在线加入内标校正干扰效应,元素检出限为0.002-0.179 μg,/L,加标回收率为96.50 %~103.10%,相对标准偏差(RSD)<2.31%.试验结果表明该法测...

关 键 词:电感耦合等离子体质谱法  地下水  痕量元素

Determination of Trace Elements in Groundwater by the Method of Inductively Coupled Plasma-Mass Spectrometry(ICP-MS)
Cai Shuwei,Cheng Qing.Determination of Trace Elements in Groundwater by the Method of Inductively Coupled Plasma-Mass Spectrometry(ICP-MS)[J].Water Purifcation Technology,2011(4):56-58.
Authors:Cai Shuwei  Cheng Qing
Affiliation:Cai Shuwei,Cheng Qing (Nanjing Institute of Geology and Mineral Resources,Nanjing 210016,China)
Abstract:Inductively coupled plasma-mass spectrometry(ICP-MS) was applied to the determination of trace elements such as Cd, Co, Cr, Cu, Mn, Mo, Ni, Pb, As, Se, Zn, Ba and Fe in groundwater and the interferential effect was corrected by the in-line addition of internal standards. The detection limits of those elements are 0.002 - 0.179 μg/L, the recovery rates are 96.50 % - 103.10 %, and the relative standard deviation (RSD) is less than 2.31%.
Keywords:inductively coupled plasma-mass spectrometry(ICP-MS) groundwater trace element  