首页 | 本学科首页   官方微博 | 高级检索  
     


Design of the high voltage isolation transmission module with low delay for ECRH system on J-TEXT
Affiliation:State Key Laboratory of Advanced Electromagnetic Engineering and Technology, Huazhong University of Science and Technology, Wuhan 430074, People's Republic of China
Abstract:As a flexible auxiliary heating method,the electron cyclotron resonance heating (ECRH) has been widely used in many tokamaks and also will be applied for the J-TEXT tokamak.To meet requirements of protection and fault analysis for the ECRH system on J-TEXT,signals of gyrotrons such as the cathode voltage and current,the anode voltage and current,etc should be transmitted to the control and data acquisition system.Considering the high voltage environment of gyrotrons,isolation transmission module based on FPGA and optical fiber communication has been designed and tested.The test results indicate that the designed module has strong anti-noise ability,low error rate and high transmission speed.The delay of the module is no more than 5 μs which can fulfill the requirements.
Keywords:J-TEXT tokamak  electron cyclotron resonance heating  high-voltage isolation  
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《等离子体科学和技术》浏览原始摘要信息
点击此处可从《等离子体科学和技术》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号