首页 | 本学科首页   官方微博 | 高级检索  
     


Suggested diagnostic reference levels for paediatric X-ray examinations in India
Authors:Sonawane A U  Sunil Kumar J V K  Singh Meghraj  Pradhan A S
Affiliation:Radiological Safety Division, Atomic Energy Regulatory Board, Mumbai, India. ausonawane@aerb.gov.in
Abstract:Exposure of children to ionising radiation is considered to carry higher risk than that of adults; therefore a need to suggest diagnostic reference levels (DRLs) for the common paediatric diagnostic X-ray procedures was recognised for the X-ray machines meeting the requirements of the recently implemented Safety Code for Medical Diagnostic X-ray Equipment and Installations in India. Measurements were carried out for entrance surface air kerma (free in air) in conventional paediatric X-ray diagnostic examinations among four age groups: <1, 1-4, 5-9 and 10-15 y. A total of 2240 air kerma measurements at different fixed focus to skin distances were studied for 7 paediatric diagnostic examinations with 11 different projections on 62 X-ray machines installed in 22 selected hospitals in the country. The third quartile values of air kerma per paediatric examination for the age group of 5-9 y were considered as values of paediatric DRLs. The suggested values of DRLs are 0.2 mGy for chest AP/PA, 0.3 mGy for chest LAT, 0.7 mGy for lumber spine AP, 1.3 mGy for lumber spine LAT, 0.3 mGy for thoracic spine AP, 0.6 mGy for thoracic spine LAT, 0.5 mGy for abdomen AP, 0.7 mGy for pelvis AP, 0.6 mGy for skull PA, 0.5 mGy for skull LAT and 0.8 mGy for hip joints AP.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号