首页 | 本学科首页   官方微博 | 高级检索  
     


TLD array for precise dose measurements in stereotactic radiation techniques
Authors:A Ertl  RF Hartl  M Zehetmayer  K Kitz  W Griffitt
Affiliation:Department of Neurosurgery, University of Vienna, Austria.
Abstract:We developed a new TLD array for precise dose measurement and verification of the spatial dose distribution in small radiation targets. It consists of a hemicylindrical, tissue-equivalent rod made of polystyrene with 17 parallel moulds for an exact positioning of each TLD. The spatial resolution of the TLD array was evaluated using the Leskell spherical phantom. Dose planning was performed with KULA 4.4 under stereotactic conditions on axial CT images. In the Leksell gamma unit the TLD array was irradiated with a maximal dose of 10 Gy with an unplugged 14 mm collimator. The doses delivered to the TLDs were rechecked by diode detector and film dosimetry and compared to the computer-generated dose profile. We found excellent agreement of our measured values, even at the critical penumbra decline. For the 14 mm and 18 mm collimator and for the 11 mm collimator combination we compared the measured and calculated data at full width at half maximum. This TLD array may be useful for phantom or tissue model studies on the spatial dose distribution in confined radiation targets as used in stereotactic radiotherapy.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号