A new use for an old Holter monitor: an ambulatory cough meter
Authors:AB Chang  RG Newman  PD Phelan  CF Robertson
Affiliation:Dept of Thoracic Medicine, University of Melbourne, Royal Children's Hospital, Victoria, Australia.
Abstract:Cough is commonly used as an outcome measure in clinical studies, although the subjective reporting of cough is unreliable when compared to objective measures. We describe an inexpensive new ambulatory cough meter that is based on a disused Holter monitor. The cough meter consists of a Holter monitor and a cough processor, designed on a computer to select the most appropriate filters. The cough meter was then validated against the overnight tape recorder on 21 occasions in 18 children (aged 6-15 yrs). The agreement between the cough meter and the tape recorder was good (mean difference of -0.3 coughs x h(-1); limits of agreement -2.2 to 1.7 coughs x h(-1)). We conclude that our newly described ambulatory cough meter provides a valid and inexpensive method of objectively monitoring cough for up to 24 h.