Photon contribution to ambient dose equivalent H(10) in the wide-spectrum neutron reference fields of the IRSN |
| |
Authors: | Langner F Löb S Nolte R Gressier V Asselineau B Lacoste V Lebreton L |
| |
Affiliation: | Physikalisch-Technische Bundesanstalt, Braunschweig, Germany. |
| |
Abstract: | The photon contribution to ambient dose equivalent in several wide-spectrum reference neutrons fields of the Institute for Radiological Protection and Nuclear Safety were measured using a Geiger-Müller counter. For the investigated fields, the ratio of photon to neutron ambient dose equivalent ranged between 0.03 and 0.20. The results show that the Geiger-Müller tube is a versatile instrument for dosimetry in mixed photon-neutron fields if sufficient information for the calculation of corrections is available. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |