首页 | 本学科首页   官方微博 | 高级检索  
     


Characterization of high-frequency, single-element focused transducers with wire target and hydrophone
Authors:Huang Bin  Shung K Kirk
Abstract:In this paper, a wire-target technique was used for lateral beam profile measurements for a single-element, focused transducers in the very high-frequency range (35-60 MHz). Two wire targets made from 9-cm long tungsten wires with diameters of 8 microm and 20 microm were used as the pulse-echo targets to measure the lateral beam profiles at the focal plane of two single-element, focused transducers, a spherically focused 40 MHz transducer and a lens-focused in-house lithium niobate (LiNbO3) 60 MHz transducer. For comparison, measurements on the same transducers were performed by three small-aperture hydrophones with geometrical diameters varying from 37 microm to 150 microm. Tomographic reconstruction of the acoustic field from the spherically focused transducer also was conducted. Results obtained with the wire-target technique are comparable to those obtained with small-aperture hydrophones in characterizing lateral radiation patterns of a single-element, focused transducer in the high-frequency range (35-60 MHz). However, the wire-target method may overestimate pulse length because of the additional attenuation caused by the return path. Compared to small-aperture hydrophones, the wire-target technique is simpler and more cost effective. Its major advantage, however, is in the frequency range above 100 MHz in which commercial hydrophones are not yet available.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号