首页 | 本学科首页   官方微博 | 高级检索  
     


RF Susceptibility of an EEG and Considerations for Attenuating RFI in Hospitals
Authors:Maskell  Stephen J
Affiliation:LBC&W/Maguire, Two Skyline Place/Suite 800, 5203 Leesburg Pike, Falls Church, VA 22041.;
Abstract:This paper discusses the testing accomplished to determine the radio frequency (RF) susceptibility of a typical electroencephalograph (EEG) and to determine the effectiveness of lead X-ray shielding as an RF shield. New data are presented on threshold field strengths and frequencies which just begin to cause interference on a typical EEG brain scan. Results are combined with data from other studies to show that a typical EEG machine would be immune to the RF environment expected in the vicinity of most hospitals. Therefore, expensive RF shielding in a hospital EEG suite is usually not required. Finally, an inexpensive shielding system utilizing X-ray shielding which was developed by the U.S. Veterans Administration is presented.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号